Table 2. configuration summary, Measuring ground-referenced signals, Cabling options – National Instruments PXI Terminal Block NI TB-2709 User Manual

Page 6

Advertising
background image

NI TB-2709 Installation Guide

6

ni.com

Measuring Ground-Referenced Signals

It is possible to set the switch on the TB-2709 to either the floating or
ground-referenced source position to measure ground-referenced signals.
However, for best results, use the OFF (ground-referenced source) switch
position to avoid ground loops. Refer to the S Series User Manual for more
information on measuring floating and ground-referenced signals.

Cabling Options

This section describes the cabling options for accessing digital signals on
the TB-2709.

A 14-pin connector on the bottom of the TB-2709 provides access to
digital I/O signals. Use the MFIT-Pigtail Cable Assembly (part number
194123-01) to access digital signals on the TB-2709. The MFIT-Pigtail
Cable Assembly features a 14-pin I/O connector on one end and pigtail

Table 2. Configuration Summary

TB-2709 Switch

Configuration

Signal Source Types for S Series Devices

Floating Source

Ground-Referenced Source

Floating Source

Ground-
Referenced Source

+

Signal

Source

TB-2709

DAQ Device

(Differential Input

Mode)

AI GND

+

Recommended

+

Signal

Source

TB-2709

DAQ Device

(Differential Input

Mode)

AI GND

+

Not Recommended

+

Signal

Source

TB-2709

Improper Configuration

DAQ Device

(Differential Input

Mode)

+

AI GND

+

Signal

Source

TB-2709

DAQ Device

(Differential Input

Mode)

AI GND

+

Recommended