NI 9237 Operating Instructions and Specifications
18
ni.com
Figure 4. Typical Passband Flatness for the NI 9237
Stopband
The filter significantly attenuates all signals above the stopband
frequency. The primary goal of the filter is to prevent aliasing.
Therefore, the stopband frequency scales precisely with the data
Frequency/Data Rate
0.5
0.4
0.025
0.000
–0.025
–0.050
0.3
0.2
0.1
0
Gain (dB)