National Instruments 24-Bit Half/Full-Bridge Analog Input Module NI 9237 User Manual

Page 22

Advertising
background image

NI 9237 Operating Instructions and Specifications

22

ni.com

Type of ADC..................................... Delta-Sigma (with analog

prefiltering)

Sampling mode ................................. Simultaneous

Internal master timebase (f

M

)

Frequency ................................... 12.8 MHz
Accuracy..................................... ±100 ppm max

Data rate range (f

s

) using internal master timebase

Minimum.................................... 1.613 kS/s
Maximum ................................... 50 kS/s

Data rate range (f

s

) using external master timebase

Minimum.................................... 390.625 S/s
Maximum ................................... 51.3 kS/s

Data rates

1

(f

s

) ...................................

, n = 1, 2, …, 31

Typical input range ........................... ±25 mV/V

Scaling coefficient ............................ 2.9802 nV/V per LSB

1

The data rate must remain within the appropriate data rate range. Refer to the

Understanding NI 9237 Data Rates

section for more information.

f

M

256

÷

n

---------------------

Advertising