Physical characteristics, Safety voltages, Safety standards – National Instruments Fieldpoint CFP-2220 User Manual

Page 33

Advertising
background image

© National Instruments Corporation

33

cFP-2200/2210/2220

Physical Characteristics

If you need to clean the controller, wipe it with a dry towel.

Screw-terminal wiring............................ Use copper conductor wire with

10 mm (0.39 in.) of insulation
from the ends.

External digital I/O terminals ......... 16 to 28 AWG

Power supply terminals................... 14 to 30 AWG

Torque for screw terminals .................... 0.5 to 0.6 N · m

(4.4 to 5.3 lb · in.)

Weight .................................................... 278 g (9.8 oz)

Safety Voltages

Connect only voltages that are within these limits.

V terminal to C terminal ........................ 35 V max, Measurement

Category I

Measurement Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS voltage. MAINS is a hazardous live electrical supply system that
powers equipment. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements include
signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics.

Caution

Do not connect the system to signals or use for measurements within

Measurement Categories II, III, or IV.

Safety Standards

This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:

IEC 61010-1, EN 61010-1

UL 61010-1, CSA 61010-1

Note

For UL and other safety certifications, refer to the product label or visit

ni.com/

certification

, search by model number or product line, and click the appropriate link

in the Certification column.