Analog output output characteristics, Dc transfer characteristics – National Instruments NI 6115/6120 User Manual

Page 91

Advertising
background image

Appendix A

Specifications

NI 6115/6120 User Manual

A-12

ni.com

Onboard calibration reference

Level ................................................5.000 V (±2.5 mV)

(actual value stored in EEPROM)

Temperature coefficient...................±4.1 ppm/°C max

Long-term stability ..........................

Analog Output

Output Characteristics

Number of channels................................2 voltage

Resolution

NI 6115............................................12 bits, 1 in 4,096

NI 6120............................................16 bits, 1 in 65,536

Max update rate

1 channel..........................................4 MS/s, system dependent

2 channel..........................................2.5 MS/s, system dependent

Output buffer size ...................................16 or 32 MS

Data transfers ..........................................DMA, interrupts,

programmed I/O

DMA modes ...........................................Scatter-gather

DC Transfer Characteristics

INL

NI 6115............................................±0.5 LSB typ, ±2 LSB max

NI 6120............................................±0.35 LSB typ, ±1 LSB max

DNL

NI 6115............................................±0.25 LSB typ, ±1 LSB max

NI 6120............................................±0.2 LSB typ, ±1 LSB max

Offset, gain error

NI 6115............................................Refer to Table A-5

NI 6120............................................Refer to Table A-6

6 ppm/ 1,000 h

±