Specifications

NI 6115/6120 User Manual

A-18

ni.com

Relative humidity ...................................10 to 90% noncondensing

Pollution Degree (indoor use only) ........2

Safety

The NI 6115/6120 was evaluated using the criteria of EN 61010-1 a-2:1995
and meets the requirements of the following standards for safety and
electrical equipment for measurement, control and laboratory use:

EN 61010-1, IEC 61010-1

UL 3111-1

CAN/CSA C22.2 No. 1010.1

Maximum Working Voltage

Maximum working voltage refers to the signal voltage plus the
common-mode voltage.

Channel-to-earth .....................................42 V, Installation Category I

Channel-to-channel.................................42 V, Installation Category I

Electromagnetic Compatibility

EMC/EMI ...............................................CE, C-Tick, and FCC Part 15

(Class A) Compliant

Emissions................................................EN 55011 Class A at 10 m

FCC Part 15A above 1 GHz

Immunity ................................................Evaluated to EN 61326:1998,

Table 1

Note

For full EMC and EMI compliance, you must operate this device with shielded

cabling. In addition, all covers and filler panels must be installed. Refer to the Declaration
of Conformity (DoC) for this product for any additional regulatory compliance
information. To obtain the DoC for this product, click Declaration of Conformity at

ni.com/hardref.nsf/

. This Web site lists the DoCs by product family. Select the

appropriate product family, followed by the product, and a link to the DoC (in Adobe
Acrobat format) appears. Click the Acrobat icon to download or read the DoC.