National Instruments Module NI PXI-8250 User Manual

Page 10

Advertising
background image

Chapter 1

Getting Started

© National Instruments Corporation

1-3

NI PXI-8250 User Manual

If you need to clean the device, use a soft, nonmetallic brush. Make sure
that the device is completely dry and free from contaminants before
returning it to service.

Operate the device only at or below Pollution Degree 2. Pollution is foreign
matter in a solid, liquid, or gaseous state that can reduce dielectric strength
or surface resistivity. The following is a description of pollution degrees:

Pollution Degree 1 means no pollution or only dry, nonconductive
pollution occurs. The pollution has no influence.

Pollution Degree 2 means that only nonconductive pollution occurs in
most cases. Occasionally, however, a temporary conductivity caused
by condensation must be expected.

Pollution Degree 3 means that conductive pollution occurs, or dry,
nonconductive pollution occurs that becomes conductive due to
condensation.

Note

The NI PXI-8250 is intended for indoor use only.

You must insulate signal connections for the maximum voltage for which
the device is rated. Do not exceed the maximum ratings for the device. Do
not install wiring while the device is live with electrical signals. Do not
remove or add connector blocks when power is connected to the system.
Remove power from signal lines before connecting them to or
disconnecting them from the device.

Operate the device at or below the installation category

1

marked on the

hardware label. Measurement circuits are subjected to working voltages

2

and transient stresses (overvoltage) from the circuit to which they are
connected during measurement or test. Installation categories establish
standard impulse withstand voltage levels that commonly occur in
electrical distribution systems. The following is a description of installation
categories:

Installation Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS

3

voltage. This category is for measurements of voltages from

specially protected secondary circuits. Such voltage measurements
include signal levels, special equipment, limited-energy parts of

1

Installation categories, also referred to as measurement categories, are defined in electrical safety standard IEC 61010-1.

2

Working voltage is the highest rms value of an AC or DC voltage that can occur across any particular insulation.

3

MAINS is defined as a hazardous live electrical supply system that powers equipment. Suitably rated measuring circuits may

be connected to the MAINS for measuring purposes.

Advertising