Wiring TEDS Channels

The NI 9219 supports only Class II TEDS sensors. Connect the two
TEDS lines to TEDS Data (T+) and TEDS COM (T–) and ensure
that neither T+ nor T– is tied in common to any of the signal inputs
(terminals 3 through 6) on the NI 9219. Visit

ni.com/info

and

enter the info code

rdteds

for information about TEDS sensors.

Grounding and Shielding Considerations

You can connect ground-referenced or floating signal sources to
the NI 9219. If you make a floating connection between the signal
source and the NI 9219, make sure the voltages on the positive and
negative connections are within the channel-to-earth working
voltage range to ensure proper operation of the NI 9219. Refer to
the

Specifications

section for more information about operating

voltages and overvoltage protection.

Note

For best signal quality, National Instruments

recommends using shielded cables and twisted pair
wiring whenever possible.

Figures 3 and 4 illustrate connecting grounded and floating signal
sources to the NI 9219 in Voltage mode.