Input impedance

Voltage and Digital In modes
(

±60 V, ±15 V, ±4 V

) ..................... 1 M

Ω

Current mode.............................. < 40

Ω

All other modes .......................... >1 G

Ω

Accuracy

Mode, Range

Gain Error

(% of

Reading)

Offset Error

(ppm of

Range)

Typ (25 °C, ±5 °C),

Max (–40 to 70 °C)

Voltage,

±

60 V

±0.3, ±0.4

±20, ±50

1

Voltage,

±

15 V

±0.3, ±0.4

±60, ±180

Voltage,

±

4 V

±0.3, ±0.4

±240, ±720

Voltage,

±

1 V

±0.1, ±0.18

±15, ±45

1

Voltage/Thermocouple,

±

125 mV

±0.1, ±0.18

±120, ±360

Current,

±

25 mA

±0.1, ±0.6

±30, ±100

4-Wire and 2-Wire

2

Resistance, 10 K

Ω

±0.1, ±0.5

±120, ±320

4-Wire and 2-Wire

2

Resistance, 1 K

Ω

±0.1, ±0.5

±1200, ±3200