Figure 5. input circuitry for one channel, R to figure 5, Voltage and current modes – National Instruments NI 9219 User Manual
Page 13
Advertising
© National Instruments Corp.
13
NI 9219 Operating Instructions
Figure 5. Input Circuitry for One Channel
Voltage and Current Modes
In Voltage and Current modes, connect the signal source to the
NI 9219 across the HI and LO terminals. The current is computed
from the voltage that the ADC measures across an internal shunt
resistor. Refer to Figure 6 for an illustration of the connections.
Figure 6. Connections in Voltage, Current, and Digital In Modes
Isolated
ADC
TEDS
1
2
+
–
CHx
3
6
NI 9219
Filter
MUX
Current mode
HI
LO
ADC
I
R
SH
NI 9219
Voltage/Digital In modes
HI
LO
ADC
V
+
–
NI 9219