Table a-2. terminal block maximum voltages, Maximum working voltage, Environmental – National Instruments Module SCXI-1125 User Manual

Page 77

Advertising
background image

Appendix

A Specifications

SCXI-1125 User Manual

A-8

ni.com

Maximum Working Voltage

Maximum voltage rating refers to the signal voltage plus the common
mode voltage (Signal + common mode). Voltage of each input shall remain
within 300 V of ground.

Channel-to-earth .....................................300 V, Measurement Category II

Channel-to-channel.................................300 V, Measurement Category II

Cautions

The SCXI-1125 is rated for Measurement Category II and is intended to carry

signal voltages no greater than 300 V. Do not use the SCXI-1125 for connection to signals
or for measurements within Categories III or IV.


When hazardous voltages (>42.4 V

pk

/60 VDC) are present on any channel, all channels are

considered hazardous. Ensure that external wiring or any circuits connected to the device
are properly insulated from human contact.

Environmental

Operating temperature ............................0 to 50

°C

Storage temperature ................................–20 to 70

°C

Humidity .................................................10 to 90% RH, noncondensing

Maximum altitude...................................2,000 meters

Pollution Degree (indoor use only) ........2

Table A-2. Terminal Block Maximum Voltages

Module

Signal Range

Maximum Voltage and Category [Insulation]

SCXI-1313A

150 V

150 V, Measurement Category II

[channel-to-channel, channel-to-earth]

SCXI-1327

300 V

300 V, Measurement Category II

[bank-to-bank, bank-to-earth]

TBX-1316

600 V

600 V, Measurement Category II

[Basic; channel-to-channel]

1000 V

1000 V, Measurement Category I

[Basic; channel-to-channel]