Appendix d common questions, Common questions – National Instruments Module SCXI-1125 User Manual

Page 84

Advertising
background image

© National Instruments Corporation

D-1

SCXI-1125 User Manual

D

Common Questions

This appendix lists common questions related to the use of the SCXI-1125.

The SCXI-1125 is backward compatible with the SCXI-1120, but what
are the major differences between the SCXI-1120 and the SCXI-1125?

Table D-1 compares the major specifications and features of the two
modules. Other specifications and features of the SCXI-1125 are the same
or very similar to the SCXI-1120.

Table D-1. Comparison of the SCXI-1125 with the SCXI-1120

Feature

SCXI-1120

SCXI-1125

Analog input

8

8

Input range

±5 V, ±250 V with SCXI-1327

±5 V, ±300 V with SCXI-1327,

±1000 VDC with TBX-1316

Isolation

250 V

rms

300 V

rms

Gains

1, 2, 5, 10, 20, 50, 100, 200, 250, 500,
1000, and 2000 jumper selectable

1, 2, 5, 10, 20, 50, 100, 200, 250, 500,
1000, and 2000 software selectable

Filters

4 Hz or 10 kHz jumper selectable

4 Hz or 10 kHz software selectable

Autozero

Not supported

Software configurable and scannable

Calibration

Manually rotateable potentiometers
for one-point offset calibration

Software internal one-point offset
calibration, software external
two-point offset and gain calibration,
and onboard calibration constant
storage in EEPROM

Scanning

333 kS/s with consecutive channels

333 kS/s with nonconsecutive and
repeating channels (random scanning)

CJC scanning

M TEMP (non-scannable) or
D TEMP (direct channel)

M TEMP, CJ TEMP (scannable)

Offset error

±6 µV ± 3 mV/gain

0.2 mV/gain typical

Gain error

±0.2% typ, ±0.6% max

±0.03% typ, ±0.08% max