Chapter 3

Signal Connections

BNC-2140 User Manual

3-6

ni.com

Caution

When you configure an analog output channel in the SE mode, the voltage

between AOGND and –DACxOUT must not exceed ±7.07 V (5 V

rms

). Voltage that exceeds

this rating can damage the BNC-2140, the DSA plug-in device, and the computer. National
Instruments is not responsible for any damages resulting from connections that exceed this
rating.

Advertising