General, Part numbers, Input power cable – Olicom 9600 User Manual

Page 14: Customer supplied), 600 vac, 10 gauge, 3 conductor wire with ground, Must conform to local electrical codes, Nominal helium pressure, Refer to table 4-1, Ambient operating temperature range

Advertising
background image

9600 Compressor Installation, Operation, and Maintenance

P/N 8040540

1-7

C

HELIX TECHNOLOGY CORPORATION

-

TI

CRYOGENICS

General

The information in Table 1-4 provides general Compressor operating
specifications.

NOTE: The 9600 Compressor is designed for continuous operation and
should remain ON when the cryopumps are in a regeneration cycle.

Table 1-4: General Compressor Operating Specifications

Specification

Values

Part Numbers

Input Power Cable

(Customer Supplied)

600 VAC
10 Gauge, 3 conductor wire with ground
Must conform to local electrical codes

Nominal Helium Pressure

Refer to Table 4-1

Ambient Operating

Temperature Range

50 - 100º F (10 - 38º C)

Interface

Gas Supply Connector

Gas Return Connector

Remote Control Receptacle

Cryopump Power Receptacles: mates with the
CTI-CRYOGENICS supplied cryopump power cable for
single pump use.

Mates with remote junction box power cable for multiple
cryopump use.

1/2 in. Aeroquip self-sealing coupling

1/2 in. Aeroquip self-sealing coupling

24VAC, 2.7A inductive mates with P5 connector P/N
MS3106A*

Adsorber Service Schedule

3 Years

* Supplied by CTI-CRYOGENICS