Helium supply, Multiple cryo-torr cryopump connections, Caution – Olicom 9600 User Manual

Page 36: Helium line connections, Multiple cryo-torr cryopump connections -13, Helium line connections -13, Splitter box located at compressor) -13, Ti cryogenics, Multiple cryo-torr cryopump connections caution

Advertising
background image

9600 Compressor Installation, Operation, and Maintenance

P/N 8040540

3-13

C

HELIX TECHNOLOGY CORPORATION

-

TI

CRYOGENICS

Figure 3-7: Alternative Multiple On-Board Cryopump or Waterpump Installation

(Splitter Box located at Compressor)

Multiple Cryo-Torr Cryopump Connections

CAUTION

Make sure the Compressor power is OFF before making any
connections to the rear panel.

Helium Line Connections

CAUTION

The use of several compressors on a single manifold feeding a common
supply header and a common return header requires special
precautions. Contact CTI-CRYOGENICS for a review of the intended
installation and for specific technical instructions.

The use of a 9600 compressor on a manifold with other
CTI-CRYOGENICS compressor models requires a reduction of the
helium charge pressure to 200 - 210 psig charge pressure to avoid
helium safety valves from inadvertently venting. Refer to Section 4 -
Operation
and Section 5 - Maintenance for more information.

HELIUM SUPPLY

HELIUM RETURN

WATER

POWER

ON-BOARD POWER CABLES

REMOTE

ON-BOARD

SPLITTER

BOX