Power cable connections, Power cable connections -14, Ti cryogenics – Olicom 9600 User Manual

Page 37

Advertising
background image

Installation

3-14

P/N 8040540

C

HELIX TECHNOLOGY CORPORATION

-

TI

CRYOGENICS

1. Connect the Supply and Return lines to the 9600 Compressor as

described in Connecting/Disconnecting Helium Flex Lines
within this section.

2. Connect the Gas Return Line to the customer supplied helium

manifold and then to the Gas Return connector on the Cryo-Torr
Cryopump.

3. Connect the Gas Supply Line to the customer supplied helium

manifold and then to the Gas Supply connector on the Cryo-Torr
Cryopump.

Power Cable Connections

1. Connect the Cryo-Torr power cable between the CRYOPUMP

ELECTRICAL OUTLET on the rear panel of the Compressor
and the CRYOPUMP ELECTRICAL INPUT on the Cryo-Torr
Interface as shown in Figures 3-8 or 3-9.

2. Connect the Cryo-Torr Power Cables between the CRYOPUMP 1,

2, or 3 connectors on the Cryo-Torr Interface and the respective
Cryo-Torr Cryopumps as shown in Figures 3-8 or 3-9.

3. Connect the User Remote cable to the Cryo-Torr Interface as shown

in Figures 3-8 or 3-9.

4. Connect the Remote cable between the Cryo-Torr Interface and the

Compressor as shown in Figures 3-8 or 3-9.

NOTE: Your installation may vary based upon the Cryo-Torr Cryopump
models used. Refer to Appendix A to consult your local
CTI-CRYOGENICS Customer Support Center for information on specific
compressor/cryopump applications.