Configuring Routing

Routing User Guide

2

On the DICOM Gateway, use a text editor to open

x:\DICOM\Dict\SCU_List.dic

, where

x

is the name of the drive used by the

DICOM Gateway to store dictionary files.

3

Add the information for each DICOM destination as shown below (comment lines are
preceded by #). Specific fields are defined in the next section.

# User Application List
# Format:
# line 1:App Name|Called AE|Calling AE|Destination IP Address|Socket
# line 2:|Presentation Context Name|Transfer Syntax Name
# line 3:||Transfer Syntax Name (if there are more than one)
#
MERGE EFILM|VistA_Storage|VistA_Send_Image|111.222.33.44|4006
|Modality Worklist Information Model - FIND|Implicit VR Little Endian
|Verification SOP Class|Implicit VR Little Endian

4

Save the file into the directory noted in Step 1.

 When using programs such as Microsoft Word, make sure that the file is saved in

“text only” mode.

 Confirm that the file extension is not changed by the text editor. The file will only

be imported if it is named

SCU_List.dic

and if it is stored in the proper

directory.

5

Open a terminal window and log into the DICOM Gateway software.

6

Enter

4

to select “System Maintenance” from the DICOM Gateway Menu

[

MAGDMENU

].

7

Enter

2

to select “Gateway Configuration and DICOM Master Files”.

8

Enter

5

to select “Update SCU_List.dic”

9

When you are prompted to update the file, enter

Y

.

Ready to read dictionary file “<x>:\DICOM\Dict\SCU_List.dic”? y// y

<Enter>

10

The file will be updated. If the operation is successful, you will be prompted to return
to the menu.

 If you encounter any errors, record the information and then press Enter. You will

need to correct the offending information and repeat the ten steps above.