INPUT:RANGE

The input range command selects the input range of one or more channels. The input range may
be either set for ±10 V or ±100 V. It is important to note that the input offset is normalized to
±10 V range. The actual input offset in the 100 V range is ten times the set value.

INPut:RANGe <range>,<channel_list>

Where <range> is 10 V | 100 V.

Where <channel_list> is the standard
channel list format supporting channels 1
through 16.

EXAMPLES


INPut:RANGe 100,(@1:16)

Sets the input range for Channels 1 through
16 to 100 V.


INPut:RANGe? 7

Returns the input range for Channel 7 as

100

100 V.


INP:RANG 10,(@4:6)

Sets the input range for Channels 4 through
6 to 10 V.


INP:RANG? 5

Returns the input range for Channel 5 as

10

10 V.