VM4016 Command Dictionary

75

INPut:OFFSet

Purpose

Sets the input threshold for a group of channels

Type

Setting

Command Syntax

INPut:OFFSet <voltage_level>,<channel_list>

Command Parameters

<voltage_level> = +9.96 V to -10.00 V.
<channel_list> = standard channel list syntax supporting channels 1 to 16.

*RST Value

0.496V for all channels

Query Syntax

INPut:OFFSet? <channel>

Query Parameters

<channel> = 1 to 16

Query Response

ASCII numeric value from -10.00 to +9.96

Description

The INPut:OFFSet command sets the input threshold for a channel or group of
channels, over which the input signal must cross to cause an interrupt event. This
command sets the value in the 8-bit DAC to which the input signal is compared. It is
important to note that the actual input offset value is affected by the INPut:RANGe
command, as the response has been normalized to ±10 V range. The actual input offset
for the allowable ranges are as follows:


Range

Entered Threshold

Actual Threshold

±10.0

x

1.0x

±100.0

x

10.0x

Examples

Command / Query

Response (Description)

INP:RANG 100,(@9:16)

(Selects an input range of ±100 V for Channels 9 -
16)

INP:OFFS 2.5,(@9:16)

(Selects an input threshold of 25 V for Channels 9
- 16)

INP:OFFS? 11

2.500 (Returns the set input threshold for Channel
11 of 25 V)

INP:RANG 10,(@1:8)

(Selects an input range of ±10 V for Channels 1 -
8)

INP:OFFS 2.5,(@1:8)

(Selects an input threshold of 2.5 V for Channels 1
- 8)

INP:OFFS? 5

2.500 (Returns the set input threshold for Channel
11 of 2.5 V)

Related Commands

INPut:RANGe <range>,<channel_list>
INPut:POLarity <polarity>,<channel_list>