Factory default settings, Error messages – Xantrex Technology XDL 35-5T User Manual

Page 24

Advertising
background image

20

Alarm output 'open' for main Output off, 'closed' for main Output on

21

Alarm output 'closed' when Aux Output is in Current Limit.
Set with control assigned to Output 1; applies to Output 1 alarm only.

22

Alarm output 'closed' when over-temperature trip occurs

23

Alarm output 'closed' when sense trip occurs

24

Alarm output 'closed' when over-current trip occurs

25

Alarm output 'closed' when over-voltage trip occurs

26

Alarm output 'closed' when any trip occurs (factory default)

30 Buzzer

off

31

Buzzer on (factory default). A single beep indicates confirmation, a double beep
indicates a wrong entry.

40

Jog digit flashes, JOG indicator only flashes when jog is 'hidden'

41

JOG indicator always flashes, except when 'hidden' (factory default)

42

JOG indicator doesn't flash, except when 'hidden' (lazy flash)

91

Loads default calibration parameters. Refer to Service Manual

92

Shows firmware version number in the display

93

Sets these # settings to their factory default

99

Enter calibration mode. Refer to Service Manual.

Factory Default Settings

The ex-factory default settings (which will apply at first power-up) are as follows:

Range: 35V/3A
Voltage: 1·000V
Current Limit: 1·000A
OVP: 40V
OCP: 5·5A
Output:

Output off; local Sense

# Settings:

00

Main Output always off at power-up

02

Aux Output always off at power up (an Output 1 # function only).

26

Alarm output 'closed' when any trip condition occurs

31

Buzzer

on

41

JOG indicator always flashes; 'lazy' flash when hidden

RS232:

9,600 Baud (XDL 35-5TP only)

Address:

11 (XDL 35-5TP only)

Error Messages

The following hardware errors are indicated by showing the appropriate error number in the display.
The OK key will flash and if pressed the error will be ignored and operation will continue as described.

Error No. Error Description

Action on pressing OK

1

Calibration constants corrupted at power-up Loads default calibration parameters

2

# functions corrupted at power-up

Loads default # settings

3

Power-down settings not correctly loaded at
power up

Loads factory default power-up
settings

Switching the instrument off with the error message showing will leave all settings unchanged.