RS232 Interface

RS232 Interface Connector

The 9-way D-type serial interface connector is located on the instrument rear panel. The pin
connections are as shown below:

Pin Name Description

1

-

No internal connection

2

TXD

Transmitted data from instrument

3

RXD

Received data to instrument

4

-

No internal connection

5 GND

Signal

ground

6

-

No internal connection

7

RXD2

Secondary received data (addressable RS232 only)

8

TXD2

Secondary transmitted data (addressable RS232 only)

9

GND

Signal ground (addressable RS232 only)

Single Instrument RS232 Connections

For single instrument remote control only pins 2, 3 and 5 are connected to the PC. However, for
correct operation links must be made in the connector at the PC end between pins 1, 4 and 6 and
between pins 7 and 8, see diagram. Pins 7 and 8 of the instrument must not be connected to the
PC, i.e. do not use a fully wired 9–way cable.


Baud Rate is set as described above in Address and Baud Rate Selection; the other parameters
are fixed as follows:

Start Bits: 1

Parity: None

Data Bits: 8

Stop Bits: 1

Addressable RS232 Connections

For addressable RS232 operation pins 7, 8 and 9 of the instrument connector are also used.
Using a simple cable assembly, a 'daisy chain' connection system between any number of
instruments, up to the maximum of 32 can be made, as shown below:

The daisy chain consists of the transmit data (TXD), receive date (RXD) and signal ground lines
only. There are no control/handshake lines. This makes XON/XOFF protocol essential and allows
the inter-connection between instruments to contain just 3 wires. The wiring of the adaptor cable
is shown below: