Leveling the balance, Installing the pan assembly, In-service cover – Ohaus AP250E ANALYTICAL Plus Electronic Balances User Manual

Page 10: Installation

Advertising
background image

10

Leveling the Balance

Leveling the Balance

Leveling the Balance

Leveling the Balance

Leveling the Balance

The balance is equipped with a level indi-
cator on the floor of the weighing chamber
and two adjustable leveling feet at the
rear. Adjust the leveling feet until the
bubble appears in the center circle of the
level indicator.

Level

Indicator

Leveling

Feet

Installing the Pan Assembly

Installing the Pan Assembly

Installing the Pan Assembly

Installing the Pan Assembly

Installing the Pan Assembly

1.

Place the shield onto the keyed bush-
ing and press it down into place. The
bent tabs must be facing upward and
the shield should be flat against the
chamber floor.

2.

Place the ring over the shield. The lip
on the bottom of the ring should fit
against the shield to keep it from
moving.

3.

Insert the pan in the load receiver
hole, matching the key on the pan
shaft with the slot in the hole. Gently
press the pan down into place.

Ring

Pan

Keyed
Bushing

Shield

In-Service Cover

In-Service Cover

In-Service Cover

In-Service Cover

In-Service Cover

Place the cover snugly over the display
unit.