Type approved/lft, Unit selection, Menus – Ohaus AP250E ANALYTICAL Plus Electronic Balances User Manual

Page 43

Advertising
background image

43

Type Approved/LFT

Type Approved/LFT

Type Approved/LFT

Type Approved/LFT

Type Approved/LFT

LFT can be set to ON or OFF. Selecting ON automatically sets the parameters shown
in the table to conform to type approved requirements. For sealing method, refer to
Type Approved Sealing section. Default setting are shown as follows:

Lockswitch Menu

Setup Locked

Print Stable Data Only

ON

Weighing Units

Weighing Units

Weighing Units

Weighing Units

Weighing Units

g

g

g

g

g

Grams

Grams

Grams

Grams

Grams

dwt

Pennyweight

ct

Carats

oz

Ounces

ozt

Troy ounces

UNIT1

Grains

t

Taels (see note)

UNIT2

Mommes

UNIT3

Custom

1. Access the

submenu.

2.

.

3.

or

.

4.

.

Unit Selection

Unit Selection

Unit Selection

Unit Selection

Unit Selection

The Unit Selection (SEL) submenu per-
mits the selection of weighing units for use
during operation. The balance can display
weights in every unit of measure listed in
table. The default setting is shown in bold
type.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE:
If Taels is enabled, see next page before
exiting the menu.

To view or change the various weighing units:

1. Access the

submenu.

2.

.

3.

or

.

4.

for next unit status.

5. Repeat steps 2 through 4 for each unit.

6.

.

MENUS