3.3 Temperature Calibration

Note: Temperature Calibration Kit is required.

ANALYZER SETUP

LOCK OUT
WEIGHT CAL

TEMP CAL

LANGUAGE
BEEPER
TIME-DATE

TEMP CAL

REMOVE PAN SUPPORT

ABORT CALIBRATION

/

TEMP CAL

INSERT CAL FIXTURE

CLOSE COVER

ABORT CALIBRATION

TEMP CAL

CALIBRATION READY

PRESS TARE/ACCEPT

ABORT CALIBRATION

TEMP CAL 100°C

CURRENT TEMP:

¾

C

TM TO CAL POINT: 15:00 MIN

ADJ CAL READING:

100

°

C

ACCEPT NEW CAL

ABORT CALIBRATION

15

TEMP CAL 100°C

CURRENT TEMP: 100

°

C

TM TO CAL POINT: 15:00 MIN

ADJ CAL READING:

100

°

C

ACCEPT NEW CAL

ABORT CALIBRATION

Adjust to thermometer reading

/

TEMP CAL 100°C

CURRENT TEMP: 100

°

C

TM TO CAL POINT: 15:00 MIN

ADJ CAL READING:

104

°

C

ACCEPT NEW CAL

ABORT CALIBRATION

ABORT CALIBRATION

TEMP CAL 100°C

CURRENT TEMP: 100

°

C

TM TO CAL POINT: 15:00 MIN

ADJ CAL READING: 104

°

C

ACCEPT NEW CAL

Adjust to thermometer reading

/

TEMP CAL 160

°

C

ABORT CALIBRATION

CURRENT TEMP: 160

°

C

TM TO CAL POINT: 0:00 MIN

ADJ CAL READING:

163

°

C

ACCEPT NEW CAL

15

TEMP CAL 160

°

C

ABORT CALIBRATION

CURRENT TEMP: 160

°

C

TM TO CAL POINT: 0:00 MIN

ADJ CAL READING:

160

°

C

ACCEPT NEW CAL

TEMP CAL 160

°

C

ABORT CALIBRATION

CURRENT TEMP: 160

°

C

TM TO CAL POINT: 0:00 MIN

ADJ CAL READING:

163

°

C

ACCEPT NEW CAL

ANALYZER SETUP

WEIGHT CAL
TEMP CAL

TIME-DATE

RS-232
PRINT
EXIT

Open the cover, allow unit to cool, remove the calibration fixture,
replace the pan support and pan.