5. Care and Maintenance

In this section, you will learn how to keep your Moisture Analyzer in good condition and how to replace
expendable parts.

5.1 Cleaning Interior/Exterior Components

To continue to obtain precise measurements, it is advisable to clean the interior components at regular
intervals. Please note the following instructions for cleaning your instrument.

Disconnect the instrument from the power supply before cleaning.

Cleaning Temperature Sensor and Protective Glass

Check the protective glass and the temperature sensor for debris
which could impede the operation. If the glass appears dirty,
clean the surface facing the compartment using a commercial
glass cleaner. If the sensor is dirty, clean using a mild cleaning
agent.

Removing Glass for Cleaning

If the inside of the glass is dirty, open the cover and remove the
four cover screws as shown.

Remove the glass holder and glass from the cover and clean with
a commercial glass cleaner on both sides.

Reassemble after cleaning.