1 sensor calibration – Soft dB Opus Suite Building Acoustics Module Airborne Sound Isolation between Rooms ASTM E 336_ISO 140-4 User Manual

Page 16

Advertising
background image

15

6.1 Sensor Calibration

The input sensitivity can be calibrated using the calibration function and a sensor calibrator. Click the

button on the Setup interface to launch the Calibration interface.

Step 1

Adjust the calibration parameters

The defaults values are:

• Averaging time:

5 s

• Frequency:

1 kHz

• Calibrator Level:

94 dB

Step 2

Install the calibrator device on the microphone

Step 3

Click START

After the average time is elapsed, the sensitivity value will update.

Step 4

Click OK to accept the sensitivity value