Soft dB Opus Suite Building Acoustics Module Airborne Sound Isolation between Rooms ASTM E 336_ISO 140-4 User Manual

Page 9

Advertising
background image

8

Step 2

Do a Level Measurement

To proceed to a measurement, select an element to measure in the table of the main panel and click

the button

to open the measurement interface.

Once in the measurement panel, click on the

button to start the average

measurement. The acquisition end by itself once the average period is reached (as set in setup) and

can be stopped at any time.

Click on the

button to accept the measurement.

• An omnidirectional should be used

• Source at least 5 m away from the partition if possible. Otherwise, in the opposite corners

of the room.

• If several source positions, the distance between positions should be at least 2 m.

• For a fixed position measurement, the averaging time should be at least 10 seconds.

• For a moving average measurement, the averaging time should be at least 30 seconds.

• The microphones must be kept more than 1 m away from the tested partition.

• If possible, the microphones should be kept more than 1 m away from all other major

extended surface or 0.5 m away otherwise.

• The microphones should be kept way from the source: >1m if room <25m³, >1.5m if room

<100m³, >2m if room >100m³.

• If fixed microphone positions is used, at least 6 positions (as least 1 m apart) should be

used for each room.

• If moving microphones is used, the microphone speed should not exceed 0.5 m/s.

Start the

Average

Accept the

measurement