Veris Technologies MSP EC and pH (1997-2008 floppy disk drive 1.02E) - Operating Instructions User Manual

Page 34

Advertising
background image

Veris Technologies

Pub. #OM MSP03-1

34

7.

You will be asked for the ID of the electrode connected to channel 1. Use the 1 and 2 keys to
change the number and 3 to confirm:


Repeat for electrode 2’s ID and press 3 to confirm.
8.

The instrument will prompt for the electrodes to be inserted into pH buffer 4 solution; Slide pH 4
buffer solution container onto electrode holder. Press 1 to continue with calibration or 2 to exit.


9.

The instrument will read the electrodes for 10 seconds, displaying the output (as it counts
seconds):


10.

After 10 seconds, the instrument will display the final pH reading and offer the options to 1)
Accept pH 4 buffer readings; 2) Redo pH 4 buffer readings; or 3) Exit pH electrode calibration
.
If the readings are satisfactory, log pH 4 reading and press 1; if the readings are suspect, press
2 to return to step 8.


11.

After accepting the pH 4 buffer readings, the Instrument will prompt for the electrodes to be
inserted into pH 7 buffer solution.

12.

Remove the pH 4 buffer solution container from the electrode holder.

13.

Rinse the electrodes and electrode holder using the manual wash or handheld wand for at least
20 seconds.

14.

Wipe the electrode holder dry; gently blot the electrode holder face and electrodes with the
edge of a paper towel.

15.

Slide the pH buffer 7 solution container onto the electrode holder.

16.

On the Instrument, press 1 to continue with calibration or 2 to exit.

17.

The instrument will read the electrodes 10 seconds, displaying the output.

Advertising