Xylem Lab 850, starting from serial number 09470315 User Manual

Page 30

Advertising
background image

Operation

Lab 850

30

ba75890e01

11/2009

Note
Here you can cancel the calibration procedure with <MODE>. This
corresponds to a two-point calibration. The value of the zero point
(Asy) is displayed for 10 seconds, then the value of the slope (Slo) is
displayed for 10 seconds. After this the meter switches to the
measuring mode.

Continue with three-

point calibration

Note
While the zero point (ASY) is being displayed, you can change the unit
of the zero point with

<

▲><▼>

.

While the slope (SLO) is being displayed, you can change the unit of
the slope with

<

▲><▼>

.

The % display refers to the Nernst slope of 59.2 mV/pH at 25° C (100
x determined slope/Nernst slope).

The unit of zero point and slope can be permanently changed in the
measurement settings (see page 45).

9

Thoroughly rinse the electrode with distilled water.

10

If necessary, set the temperature of the third buffer solution
with <

▲><▼>.

11

Immerse the pH electrode in the third buffer solution.

12

Press the <OK>

key.

The SC display indicator flashes.
Depending on the setting, the display shows the nominal pH
value for the recognized buffer or the electrode voltage (mV).
As soon as a stable value is recognized, the value of the zero
point (ASY) is displayed for 10 seconds, then the value of the
slope (SLO) is displayed for 10 seconds.
After this the meter switches to the measuring mode.

Advertising
This manual is related to the following products: