Interval for automatic storing (AutoStore)

After setting the interval for automatic storing the current measurement
dataset is stored at the specified interval.

AutoStore

interval

AutoStore interval (Int.1)

OFF, 5 s, 10 s, 20 s, 30 s, 1 min, 2 min,
5 min, 10 min, 20 min, 30 min, 60 min

1

Press <STO

_

> to open the setting for the storage interval.

Int.1, the setting of the storage interval is displayed.

2

Press <

▲><▼>

to select an interval.

3

Confirm with <OK>.
FrEE, the number of free memory locations is displayed.

4

Confirm with <OK>.
The setting of the storage interval is completed.
The meter switches to the measuring mode.