What to do if...

55

ba75890e01

11/2009

6

What to do if...

Error message

Err1

Error message

Err2

Error message

Err4

Error message

CalError

Cause

Remedy

pH electrode:

– Measured value outside the

measuring range

– Use a suitable electrode

– Air bubble in front of the

junction

– Remove air bubble

– Air in the junction

– Extract air or moisten

junction

– Cable broken

– Replace the electrode

– Gel electrolyte dried out

– Replace the electrode

Cause

Remedy

– No electrode connected

– Connect the electrode

– Setting time during calibration

too long

– Adjust temperature if

necessary

– Recalibrate

Cause

Remedy

– Temperature not stable during

calibration.

– Adjust temperature if

necessary

– Recalibrate

Cause

Remedy

pH electrode:

– The values determined for zero

point and slope of the electrode
are outside the allowed limits.

– Recalibrate

– Junction contaminated

– Clean junction

– Electrode broken

– Replace the electrode

Buffer solutions: