7.2

Measuring ranges, resolution, accuracy

Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the electrodes and buffer solutions has to be taken into
account additionally.

Measuring ranges,

resolution

Variable

Measuring range

Resolution

pH

- 2.000 ... + 19.999
- 2.00 ... + 19.99

0.001
0.01

U [mV]

- 999.9 ... + 999.9
- 2000 ... + 2000

0.1
1

T [°C]

- 5.0 ... + 120.0

0.1

T [°F]

+ 23.0 ... + 248.0

0.1

Manual

temperature input

Variable

Range

Increment

T

manual

[°C]

- 25 ... + 125

1

T

manual

[°F]

- 13 ... + 257

1

Accuracy (± 1 digit)

Variable

Accuracy

Temperature of the test
sample

pH / range *

- 2.000 ... + 19.999

± 0.005

+ 15 °C ... + 35 °C

- 2.00 ... + 19.99

± 0.01

+ 15 °C ... + 35 °C

U [mV] / range

- 999.9 ... + 999.9

± 0.3

+ 15 °C ... + 35 °C

- 2000 ... + 2000

± 1

+ 15 °C ... + 35 °C

T [°C] / temperature sensor

NTC 30

± 0.1

PT 1000

± 0.3

* when measuring in a range of ± 2 pH around a calibration point