1 measuring the ph value, Measuring the ph value – Xylem Lab 860 User Manual

Page 20

Advertising
background image

Operation

Lab 860

20

ba75554e05

01/2010

4.3.1

Measuring the pH value

Stability control SC

(drift control)

The stability control function (drift control) checks the stability of the
measurement signal. The stability has a considerable effect on the
reproducibility of the measured value.

For identical measurement conditions, the following criteria apply:

Drift within 15 sec < 0.02 pH units.

Note
The current measurement with stability control can be terminated at
any time (accepting the current value) by pressing <OK>.

1

Perform the preparatory activities according to page 19.

2

Immerse the pH electrode in the test sample.

3

Using <MODE>, scroll as necessary until the measured
parameter pH is displayed.

8

4

2

2

9

9

6

pH

°C

TP

1

If necessary, call up the measured variable pH with <MODE>.

2

With <SC>, activate the stability control function.
The SC function display indicator appears.
The current measured value is frozen (hold function).

3

Start measurement with stability control with <OK>.
SC flashes until a stable measured value is reached.
This measured value is downloaded to the interface.

4

If necessary, start the next measurement with stability control
with <OK>.

5

To terminate the stability control function: Press the <MODE>
or <SC>

key.

Advertising
This manual is related to the following products: