Xylem Lab 860 User Manual

Page 57

Advertising
background image

Lab 860

What to do if...

57

ba75554e05

01/2010

Display

to

Obviously incorrect

measured values

Instrument does not

react to keystroke

You want to know which

software

version is in the

instrument

Message StoFull

Cause

Remedy

– Time-out of the interface

– Check that the instrument is

connected

Cause

Remedy

pH electrode:

– pH electrode unsuitable

– Use a suitable electrode

– Temperature difference

between buffer and test sample
too high

– Adjust temperature of buffer

or sample solutions

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press and hold the <SC>
key and switch the meter on

Cause

Remedy

– E. g., a question by the service

department

– Switch on the meter.

During the display test,
display the software version
with <OK>.

Cause

Remedy

– All 800 memory locations are

full

– Output data memory and

clear data memory