2 measurement settings, Measurement settings – Xylem Lab 860 User Manual

Page 46

Advertising
background image

Operation

Lab 860

46

ba75554e05

01/2010

4.7.2

Measurement settings

These settings apply to calibration and measurement (the default
setting is printed in bold).

Buffer set for pH

calibration

(SET)

Number of the buffer set for pH
calibration (SET)

1 ... 16

Display during calibration (BUFFER)

pH (buffer nominal value),
U (electrode voltage)

Unit of the value for the slope (SLO)

%, mV/pH

Unit of the value for the zero point
(ASY)

pH, mV

Resolution (rES)

HI (0.001), Lo (0.01)

Temperature unit (UnIt)

°C, °F

Calibration interval (Int.C)

0 ... 7 ... 999 d

1

Open the menu for measurement settings with <MODE

_

>.

Set 1 ... 16, the specified buffer set is displayed.

2

Using <

▲><▼>,

select a buffer set Set 1 ... 16.

3

Confirm with <OK>.
bUFF, the setting of the display during calibration is displayed.

1

AutoCal

SET