Appendix: Firmware update

General information

With the Update_Labxxx_MxxxP program and a PC you can update
the firmware of the Lab 860 to the newest version.

For the update you can connect the meter to a PC via the USB or
RS232 interface.

For the update via the USB interface, the following is required:

z a free USB interface (virtual COM port) on your PC
z the driver for the USB interface

(installation see page 41)

z the Z875 USB cable (included in the scope of delivery of the

Lab 860).

For the update via the RS232 interface, the following is required:

z a free RS232 interface on your PC
z the RS232 cable, Z390.

Program installation

With the installation program,
"Install_Update_Labxxx_MxxxP_Vx_yy_English.exe", install the
firmware update program on your PC.

Program start

Start the "Update_Labxxx_MxxxP" program from the Windows start
menu.

You can change the language via the language menu.

Firmware update

Proceed as follows:

1

Connect the Lab 860 to a USB interface (virtual COM port) of
the PC with the aid of the USB interface cable Z875.
or
Connect the Lab 860 to a serial interface (COM port) of the PC
with the aid of the interface cable Z390.

2

Make sure the Lab 860 is switched on.

3

To start the update process click the OK button.
The program automatically recognizes the used interface.