Operation

Lab 870

26

ba75555e04

11/2009

4.4

Calibration

Why calibrate?

pH electrodes age. This changes the zero point (asymmetry) and slope
of the pH electrode. As a result, an inexact measured value is
displayed. Calibration determines the current values of the zero point
and slope of the electrode and stores them. Thus, you should calibrate
at regular intervals.
For non-ID sensors, the calibration data is stored in the meter. For ID
sensors, the calibration data is stored in the sensor.

When do you have to

calibrate?

z After connecting another electrode

z If the CalClock has expired and flashes

Buffer sets for

calibration

You can use the buffer sets quoted in the table for an automatic
calibration. The pH values are valid for the specified temperature
values. The temperature dependence of the pH values is taken into
account during calibration.

No.

Buffer set*

pH values

at

1

SI Analytics GmbH DIN buffers
according to DIN 19266/NBS

1.679
4.006
6.865
9.180

12.454

25 °C

2

SI Analytics GmbH Technical
buffers according to DIN 19267

2.000
4.010
7.000

10.011

25 °C

3

Merck1*

4.000
7.000
9.000

20°C

4

Merck2 *

1.000
6.000
8.000

13.000

20°C

5

Merck3 *

4.660
6.880
9.220

20°C

6

DIN 19267 *

1.090
4.650
6.790
9.230

25 °C