3 downloading calibration data, Downloading calibration data, La c – Xylem Lab 870 User Manual

Page 35

Advertising
background image

Lab 870

Operation

35

ba75555e04

11/2009

Note
While the zero point (ASY) is being displayed, you can change the unit
of the zero point with

<

▲><▼>

.

While the slope (SLO) is being displayed, you can change the unit of
the slope with

<

▲><▼>

.

The % display refers to the Nernst slope of 59.2 mV/pH at 25° C (100
x determined slope/Nernst slope).

The unit of zero point and slope can be permanently changed in the
measurement settings (see page 44).

4.4.3

Downloading calibration data

You can download calibration data:

z to the display

– via the memory menu

– via the calibration menu

z to the interface (see page 38)

Download to display via

memory menu

While the calibration data is being displayed, you can:

1

Open the memory menu with <RCL>.

2

If necessary, scroll with <RCL> until CAL diSP is displayed.

3

Press <OK> to display the calibration data.
The following data is displayed consecutively for 10 seconds
each:
date, zero point, slope.

z

press <OK>

to display further calibration data
(date, zero point, slope)

L

A

C

9

3 0

0

Day.Month

AutoCal