3 downloading calibration data, 4 rs232 interface, Downloading calibration data – Xylem Lab 870 User Manual

Page 39: Rs232 interface

Advertising
background image

Lab 870

Operation

39

ba75555e04

11/2009

4.5.2

Automatically downloading measurement datasets at
intervals

In order to automatically download to the interface measured values at
certain time intervals, set the download interval (Int.2).

Setting the download

interval

The default setting for the download interval (Int 2) is OFF.
To switch the function on, set an interval (5 s, 10 s, 20 s, 30 s, 1 min, 2
min, 5 min, 10 min, 20 min, 30 min, 60 min):

4.5.3

Downloading calibration data

4.5.4

RS232 interface

1

Press <SC

_

> to open the setting of the Int.2 interval.

2

If necessary, set an interval with <

▲><▼>.

3

Close the setting with <OK>.
The download to the interface takes place at the specified
interval.

1

Open the memory menu with <RCL>.

2

If necessary, scroll with <RCL> until CAL Prt is displayed.

3

Press <OK> to download the calibration data to the interface.

1

Connect the interface to the PC or printer via the cable Z390
(PC) or Z393 (ext. printer).

2

If necessary, disconnect a connected USB cable from the
meter.

3

Set up the following transmission data at the PC/printer:
– Baud rate: selectable from 1200, 2400, 4800, 9600,

– Handshake: RTS/CTS + Xon/Xoff

– PC only:

– Parity: none

– Data bits: 8

– Stop bits: 2