Xylem Lab 870 User Manual
Page 30
Advertising

Operation
Lab 870
30
ba75555e04
11/2009
Preparatory activities
Clean the electrode according to the
sensor operating manual
CalError
CalError
< -30 or
> 30
... -62 or
... -50
Eliminate the error according to
page 53
CalClock
Calibra-
tion record
Zero point
[mV]
Slope
[mV/pH]
1
Switch on the meter with <On/Off>.
2
Connect a pH electrode to the meter.
3
Keep the buffer solutions ready.
4
Adjust the temperature of the solutions and measure the
current temperature if the measurement is made without
temperature sensor.
5
Set the buffer set to be used for calibration as necessary.