2 measurement settings, Measurement settings – Xylem Lab 870 User Manual

Page 44

Advertising
background image

Operation

Lab 870

44

ba75555e04

11/2009

4.6.2

Measurement settings

These settings apply to calibration and measurement (the default
setting is printed in bold).

Buffer set for pH

calibration

(SET)

Number of the buffer set for pH
calibration (SET)

1 ... 16

Display during calibration (BUFFER)

pH (buffer nominal value),
U (electrode voltage)

Unit of the value for the slope (SLO)

%, mV/pH

Unit of the value for the zero point
(ASY)

pH, mV

Resolution (rES)

HI (0.001), Lo (0.01)

Temperature unit (UnIt)

°C, °F

Calibration interval (Int.C)

0 ... 7 ... 999 d

1

Open the menu for measurement settings with <MODE

_

>.

Set 1 ... 16, the specified buffer set is displayed.

2

Using <

▲><▼>,

select a buffer set Set 1 ... 16.

3

Confirm with <OK>.
bUFF, the setting of the display during calibration is displayed.

1

AutoCal

SET