background image

Commissioning

IFL 70x IQ

3 - 6

ba76129e01

10/2012

3.4

Initial commissioning

Prerequisite for the display of measured values of the
IFL 70x IQ on a terminal of the

IQ S

ENSOR

N

ET

system is

the current software version of the controller and the termi-
nal.
The current software is available on the Internet under
www.YSI.com.

1

Install the sensor at the measuring location and establish the
connection to the

IQ S

ENSOR

N

ET

. (see section 3.3).

2

Using <>, select the IFL 70x IQ sensor in the measured
value display.

3

Open the Display/Options / Extended sensor functions menu.
The display shows the echo profile and certain special menus.

4

Open the Sensor settings / menu.

5

Adjust the following settings for the measuring location
(Sensor settings):

Immersion depth

Extended settings / Temperature
(average water temperature at mean water depth)

Water depth
(water depth to the bottom of the basin at the measuring
location)

Setting of

Temperature

If the average water temperature fluctuates due to sea-
sonal changes we recommend to adjust the mean temper-
ature on the instrument according to the current season.

Setting of Water depth
The Water depth setting should be consistent with reality as
much as possible. Therefore, we recommend to determine
the water depth at the measuring location and enter it.

6

Using Save and quit, confirm the settings and switch to the dis-
play of the echo profile.