3 recommended fields of application, Recommended fields of application -3 – YSI IFL 700 IQ User Manual

Page 7

Advertising
background image

IFL 70x IQ

Overview

1 - 3

ba76129e01

10/2012

Fig. 1-3

Principle of the ultrasonic measurement

Wiper (cleaning system)

The IFL 700 IQ sensor has a mechanical wiper that effectively cleans
gas bubbles and dirt off the ultrasonic transducer. The wiper operates
contactless and is maintenance-free and wear-free.

1.3

Recommended fields of application

Sludge level control and monitoring in waste water treatment.

1

Transmitted ultrasonic waves

2

Echo reflected by the sludge blanket (short reflection interval)

3

Echo reflected by the bottom area (long reflection interval)

Sludge blanket level

1

3

2

Ultrasonic

transducer

Bottom

Detailed information on the subject of sludge level mea-
surement is given for example in the DWA information
sheet no. 256 "Prozessmesstechnik auf Kläranlagen, Teil
8: Messeinrichtungen zur Bestimmung des Schlammspie-
gels" (Process measuring technique at wastewater treat-
ment plants, part 8: Instrumentation for determination of the
sludge level".

Advertising