IFL 70x IQ

Commissioning

3 - 9

ba76129e01

10/2012

3.5

Setting table for the IFL 70x IQ

3.5.1 Sensor settings menu

Carrying out settings

The sensor settings can be accessed from the following menus:

Using <S>, switch from the measured value display to the main
menu of the settings. Then navigate to the setting menu (setting
table) of the sensor. The exact procedure is given in the relevant
IQ S

ENSOR

N

ET

system operating manual.

Using <>, select the IFL 70x IQ sensor in the measured value
display. Using <M>, open the Display/Options / Extended sensor
functions
menu. Open the Sensor settings menu.

Default values are marked in bold.

Setting

Selection/values

Explanation

Measuring mode

Details see Fig. 3-1 in section 3.3.2

Sludge level height

Position of the sludge blanket level in rela-
tion to the bottom of the basin (SLH).

Sludge level depth

Position of the sludge blanket level in rela-
tion to the surface of the water (SLD).

Unit

m
ft

Selection of the unit for the distance
Meter
Foot

Immersion depth

0.05 ... 0.10 ... 3.00 m

Distance between the surface of the ultra-
sonic transducer (underside of the sen-
sor) and the surface of the water (see Fig.
3-1 in section 3.3.2).

Water depth

0 ... 6 ... 18 m

Vertical distance between the surface of
the water and the bottom of the basin at
the measuring location (see Fig. 3-1 in
section 3.3.2).
The water depth can be determined by
plumbing.