YSI IFL 700 IQ User Manual

Page 20

Advertising
background image

Commissioning

IFL 70x IQ

3 - 10

ba76129e01

10/2012

Setting

Selection/values

Explanation

Method

The rising side of the echo is evaluated for
the measured value determination. One of
two methods can be selected for this.

Rel. threshold

The measured value is equivalent to the
point where the intensity of the echo
reaches the adjusted relative threshold.
The value relates to the maximum of the
echo intensity (100 %):

Maximum gradient

The measured value is equivalent to the
point with the maximum gradient:

Rel. threshold value

25 ... 75 ... 100 %

Threshold value for evaluation according
to the Rel. threshold method.

100 %

Depth

Measured value

25 %

Relative
threshold

Depth

Measured value

Maximum
gradient