IFL 70x IQ

Commissioning

3 - 5

ba76129e01

10/2012

NOTE (only IFL 700 IQ)
If the sensor is connected to the IQ S

ENSOR

N

ET

, the mechanical wiper

may start moving unexpectedly. Make sure that the swivel range of the
wiper is always free. Only put the sensor down its leg supports on a flat
surface with the ultrasonic transducer pointing downward.

Fig. 3-2

Connecting the sensor

2

Plug the socket of the SACIQ (SW) sensor connection cable
onto the plug head connector of the sensor. At the same time,
rotate the socket so that the pin in the plug head connector (1)
clicks into one of the two holes in the socket.

3

Then screw the coupling ring (2) of the sensor connection
cable onto the sensor up to the stop.

SACIQ (SW)

1

2