4 measuring, Measuring -1, 4measuring – YSI IFL 700 IQ User Manual

Page 25

Advertising
background image

IFL 70x IQ

Measuring

4 - 1

ba76129e01

10/2012

4

Measuring

Factors affecting the

measured value

The following factors have an impact on the measured value:

The environmental conditions at the measuring location deviate too
much from the sensor settings (Immersion depth, Water depth,
Temperature)

The distance between the ultrasonic transducer and the sludge level
is too small (dead zone, see also section 3.3.3)

Moving fixtures cross the measuring cone or the installation location
of the sensor

Foreign bodies or air bubbles are in front of or on the ultrasonic
transducer.

1

Submerse the sensor in the sample.

2

Read the measured value on the terminal of the IQ S

ENSOR

N

ET

system.