IFL 70x IQ

What to do if...

6 - 1

ba76129e01

10/2012

6

What to do if...

Mechanical damage to

the sensor

Display "----"

(no valid measured

value)

Cause

Remedy

– Return the sensor

Cause

Remedy

Sensor is permanently in the air

Immerse the sensor in water
(see section 3.3.2)

There are too many air bubbles in
the water or on the ultrasonic trans-
ducer

Select a measurement location
free of air bubbles

Sensor is dirty

– Clean the sensor and/or its

environment

– Check the function of the

wiper (see section 3.5)

Thread algae floating in front of the
sensor

Remove the thread algae from
the sensor or its environment

In the selected Evaluation range,
no echo is available that meets all
settings

– Check whether there is a

sludge blanket in the
selected Evaluation range

– Check whether the settings

are suitable for the applica-
tion.

Defective sensor

Contact the service department