3 calibration, 1 overview, 3 calibration -4 – YSI CarboVis 705 IQ User Manual

Page 28: Overview -4, Section, Alibration

Advertising
background image

Measuring / Operation

CarboVis 70x IQ

4 - 4

ba75950e02

11/2012

4.3

Calibration

4.3.1 Overview

In the delivery condition, the sensor is precalibrated and ready to mea-

sure (see section 3.4.1).
There are two types of calibration that are included into the calculation

of the measured value at different points (see section 4.1):

User calibration (see section 4.3.2)

Sensor check/

Zero adjustment

(see section 4.3.3).

An overview diagram of the calibration procedures can be found on

page 4 - 10.

User calibration

With a user calibration you compensate for the influence of the test

sample. The user calibration requires a successful

Sensor check (H2O

dest.)

and a valid

Zero adjustment

.

We recommend to check the current user calibration against indepen-

dent comparison measurements if you suspect a basic change of the

characteristics of the test sample. If the comparison measurements

show clear deviations, these can be compensated by a user calibration.

Recalibration after

changing the measuring

location

The data of the user calibration are stored in the IQ S

ENSOR

N

ET

con-

troller and thus assigned to the measuring location (not to the sensor).

The calibration data of the

Zero adjustment

are stored in the sensor.

Each time after the application or measurement location

(and the setting,

Measuring location

) was changed, we

recommend to carry out a

Sensor check (H2O dest.)

(and, if the

Sensor check (H2O dest.)

was not successful,

an additional

Zero adjustment

if necessary) with a subse-

quent user calibration.

Advertising
This manual is related to the following products: