YSI CarboVis 705 IQ User Manual
Page 20
Commissioning
CarboVis 70x IQ
3 - 10
ba75950e02
11/2012
Measuring mode
COD spectr., total
COD spectr., dissolv
TOC spectr., correl.
BOD spectr., correl.
DOC spectr., correl.
SAC 254, total
SAC 254, dissolved
UVT 254
The measured parameter is displayed
in the selected citation form.
When changing the
Measuring mode
setting, always check the
Measuring
location
setting.
If the previously set measurement loca-
tion is not available for the selected
measuring mode, the measurement
location is automatically changed..
Measuring location
Outlet
Inlet
Activation
CarboVis 701 IQ
:
Activation
Inlet
Outlet
CarboVis 705 IQ
:
Outlet
(permanently set)
Measurement location or application of
the sensor.
The possible measurement loca-
tions are displayed depending on the
currently set measuring mode.
The sample matrix on which the cal-
culation of the measured value is
based changes with the location of
the sensor.
The measurement location
Activation
can only be selected for
Measuring
mode
COD spectr., dissolv
.
When the measurement location is
changed, all virtual sensors are
deleted.
Measuring range
-
Display of the measuring range
Cal - # raw value 1
Value pairs of the user calibration (see section 4.3.2).
Note: Leave those values unchanged during the initial commission-
ing.
Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Number of N-sensors
Number of C-sensors
1 ... 2
10 ... 43
Apart from the main sensor, up to 4
more sensors can be displayed.
The settings for the additional sensors
are done in the menus for the respec-
tive sensors.
Zero adjustment
Factory
User
Setting of which zero adjustment is
used in the sensor.
Factory zero adjustment or
Zero adjustment carried out last
Menu item
Settings
Explanations